A Fast & Reliable Liquid Chromatography Mass Spectrometry Method for Determination of Amino acids & Acylcarnitines including Succinyl acetone in dried blood spots for newborn screening.

International Journal of Medical Science
© 2019 by SSRG - IJMS Journal
Volume 6 Issue 7
Year of Publication : 2019
Authors : Shrimant N Panaskar, Mahesh H Hampe, Mr. Susheel Singh, Akshata Shivalkar, Pallavi Kadam
pdf
How to Cite?

Shrimant N Panaskar, Mahesh H Hampe, Mr. Susheel Singh, Akshata Shivalkar, Pallavi Kadam, "A Fast & Reliable Liquid Chromatography Mass Spectrometry Method for Determination of Amino acids & Acylcarnitines including Succinyl acetone in dried blood spots for newborn screening.," SSRG International Journal of Medical Science, vol. 6,  no. 7, pp. 1-11, 2019. Crossref, https://doi.org/10.14445/23939117/IJMS-V6I7P101

Abstract:

Introduction:We developed & validate first time TMS based non derivatized 1.5 min protocol to quantify 25 amino acids & 30 acylcarnitines along with Succinyl acetone.
Method: 3.2mm DBS disk was extracted with a mixture of methanol: Hydrate hydrazine: oxalic acid: Acetonitrile: Water. Analysis was performed on LCMSMS with direct injection method. Water: formic acid & Methanol was used in gradient elution. Optimized multiple reaction monitoring (MRM) was used for detection of amino acids and acylcarnitines on a Shimatzu mass spectrometer. The method was validated with respect to linearity, precision, recovery, accuracy.
Result: Precision, accuracy and analytical measurement range considering overall precision within 10% and accuracy within 15% of for all amino acids for Acylcarntine at one fortified level, which quantitated 21.8% lower than the expected value. Inter method comparison shows only 8.5% & 10.5 % error bias for AA & AC respectively. Analytes were stable for 1 month at room temperature & 6 months at -200C storage condition.
Conclusion: The present design protocol is sensitive with significant LOD and cost effective which can be effectively applied as fast primary & second-tier test for newborn screening of metabolic disorders associated with abnormal levels of acylcarnitines and amino acids, potentially reducing false positive cases and shortening the time to diagnosis.

Keywords:

Amino acids & Acylcarnitines, Dried blood spot, ESI-MS, newborn screening

References:

[1] T. Ozben, Expanded newborn screening and confirmatory follow-up testing for inborn errors of metabolism detected by tandem mass spectrometry. Clin Chem Lab Med 2013; 51:157-76.
[2] AE. Garrod, The incidence of alkaptonuria: a study in chemical individuality. 1902. Mol Med 1996; 2:274-282.
[3] P. Monique, VS. Christine, P. Konstantinos, E. Claire, S. Jean-Paul, M. Aymeric, B. Denis, ESI-MS/MS analysis of underivatised amino acids: a new tool for the diagnosis of inherited disorders of amino acids metabolism. Fragmentation study of 79 molecules of biological interest in positive and negative ionization mode. Rapid Commun. Mass Spectrom 2003; 17:1297-1311.
[4] MP. Frank, RW. Powers, Simple and rapid quantitative high-performance liquid chromatographic analysis of plasma amino acids. Chromatogr B Analyt Technol Biomed Life Sci. 2007;852:646-9.
[5] SV. Suresh Babu, MM. Shareef, PK. Shetty, KT. Shetty, HPLC method for amino acids profile in biological fluids and Inborn Metabolic Disorders of Aminoacidopathies. Indian J Clin Biochem 2002; 17:7-26.
[6] MH. Hampe, SN. Panaskar, AA. Yadav, PW. Ingale, Gas chromatography/mass spectrometry-based urine metabolome study in children for inborn errors of metabolism: An Indian experience. Clin Biochem 2017;50: 121-126.
[7] MH. Hampe, PW. Ingale, SN. Panaskar, AA. Yadav, Two tier analysis of organic acid disorders: a comprehensive approach for newborn screening. International Journal of Biomedical and Advance Research. 2015; 6: 84-90.
[8] L. Ruiting, L. Peifang, L. Peijia, T. Yuan, H. Yunfei, G. Yiqiao H. , Hua, C. Jiaqing, Z. Zunjian, H. Yin, A novel liquid chromatography tandem mass spectrometry method for simultaneous determination of branched‑chain amino acids and branched‑chain α‑keto acids in human plasma. Amino Acids 2016; 48:1523-1532.
[9] W. Chunyan, Z. Hongbin, P. Zifeng, S. Fengrui, L. Zhiqiang, L Shuying,. Classification of type 2 diabetes rats based on urine amino acids metabolic profiling by liquid chromatography coupled with tandem mass spectrometry. J.Chromatogr. B, 2013;935:26-31.
[10] DH. Chace, TA. Kalas, EW. Naylor, Use of tandem mass spectrometry for multianalyte screening of dried blood specimens from newborns. Clin Chem 2003;49:1797-817.
[11] KH. Carpenter, V. Wiley, Application of tandem mass spectrometry to biochemical genetics and newborn screening. Clin Chim Acta 2002; 322:1-10.
[12] DW. Johnson, Inaccurate measurement of free carnitine by electrospray tandem mass spectrometry screening method of blood spots. J Inherit Metab Dis 1999; 22:201-2.
[13] T. Kuster, T. Torresani, P. Kleinert, S. Durka, F. Neuheiser, CW. Heizmann, H. Troxler, Filter paper cards contaminated with EMLA cream produce artifacts on acylcarnitine analysis. J Inherit Metab Dis 2004; 27: 707-9.
[14] S.Malvagia, G. La Marca, B. Casetta, S. Gasperini, E. Pasquini, MA. Donati, E. Zammarchi, Falsely Elevated C4-carnitine as expression of glutamate formiminotransferase deficiency in tandem mass spectrometry newborn screening. J Mass Spectrom 2006; 41:263-5.
[15] JE. Abdenur, NA. Chamoles, AE. Guinle, AB. Schenone, ANJ.Fuertes. Diagnosis of isovaleric academia by tandem mass spectrometry: false positive result due to pivaloylcarnitine in a newborn screening programme. J Inherit Metab Dis 1998; 21:624-30.
[16] N. Napolitano, V. Wiley, JJ. Pitt, Pseudo-glutarylcarnitinaemia in medium-chain acyl-CoA dehydrogenase deficiency detected by tandem mass spectrometry newborn screening. J Inherit Metab Dis 2004;27:465-71.
[17] HZ .Thomas, FF. Eileen, M. Deborah, A. Cecilia, ES. Vivian, MJ. Donna, WS. Arnold,Tandem Mass Spectrometric Analysis for Amino, Organic, and Fatty Acid Disorders in Newborn Dried Blood Spots: A Two-Year Summary from the New England Newborn Screening Program. Clin Chem 2001:47:1945-1955.
[18] JV.Mei, JR .Alexander, BW.Adam, WH .Hannon. Use of filter paper for the collection and analysis of human whole blood specimens. J Nutr. 2001; 131:1631-6.
[19] AH .Christopher, RD .Víctor. Simultaneous quantitation of hexacosanoyl lysophosphatidylcholine, amino acids, acylcarnitines, and succinylacetone during FIA–ESI–MS/MS analysis of dried blood spot extracts for newborn screening. Clin Biochem. 2016; 49:161-165.
[20] NCCLS approved Guidelines EP5 –A2, Second Edition, Evaluation of Precision performance of Quantitative measurement methods.2004; NCCLS, Wayne, PA.
[21] NCCLS Approved guidelines EP6 – A, Evaluation of Quantitative measurement procedures: A Statistical Approach.2003; NCCLS, Wayne, PA.
[22] NCCLS Approved guidelines EP17 – A, Protocol for determination of Limit of detection and Limits of quantification. 2004; NCCLS, Wayne, PA.
[23] EH .Smith, D Matern. Acylcarnitine analysis by tandem mass spectrometry. Curr Protoc Hum Genet. 2010; 17:1-20.
[24] JH. Osorio, M. Pourfarzam Hydrolysis of acylcarnitines during measurement in blood and plasma by tandem mass spectrometry. Acta Bioquim Clin L. 2010; 44:189-193.
[25] MS. Watson, MA. Lloyd-Puryear, MY .Mann, P. Rinaldo, RR .Howell. Newborn screening: toward a uniform screening panel and system. Genet Med 2006; 8:1-252.
[26] A. Gucciardi, P .Pirillo, IM .Di Gangi, M. Naturale, G. Giordano A rapid UPLC-MS/MS method for simultaneous separation of 48 acylcarnitines in dried blood spots and plasma useful as a second-tier test for expanded newborn screening. Anal. Bioanal. Chem. 2012; 404:741-751.
[27] M. Peng, X. Fang, Y. Huang, Y. Cai, C. Liang, R. Lin, L. Liu, Separation and identification of underivatized plasma acylcarnitine isomers using liquid chromatography-tandem mass spectrometry for the differential diagnosis of organic acidemias and fatty acid oxidation defects. J. Chromatogr. A 2013; 1319:97-106.
[28] MS .Rashed, MP .Bucknall, D. Little, A. Awad, M. Jacob, M. Alamoudi, M. Alwattar, PT. Ozand, Screening blood spots for inborn errors of metabolism by electrospray tandem mass spectrometry with a microplate batch process and a computer algorithm for automated flagging of abnormal profiles. Clin Chem. 1997; 43:1129-41.